1. In ultrasonic diagnosis and treatment operations, it is used together with intracavitary or extracorporeal ultrasonic probe to guide and fix biopsy needle guide and other puncture instruments.

2. Open the clamp, align the puncture rack and the positioning mark of the probe, and press it to place it on the probe, and lock the upper and lower clamps

(or elastic fastening) nature.

3. After checking that the biopsy needle guide has been firmly installed, the biopsy needle guide can be inserted into the guiding structure for puncture operation.Please refer to the instruction manual or installation drawing.


Suitable for Philips C8-4v \ C9-3v \ C10-3v ultrasonic probes

1. Soak the biopsy needle guide for at least five 

    minutes in a low-foam enzyme cleaner with neutral pH.

2. When soaking, use the instrument brush to remove

    residual contaminants on the surface and parts.

    If visible contaminants are not easily removed,and 

    repeat soaking procedure.

3. Remove the biopsy needle guide from the cleaning 

    solution and use it with a rag to wipe off all residues.

4. Can be steamed (autoclave) at 132-138℃ for

    15 to 30 minutes,Sterilize or sterilize the biopsy

     needle guide.


Outpatient service, stomatological hospital, medical beauty institution, social health hospital
Usage scenario

5 / F, building A2, Mingjun Industrial Park, Xinshi community, Dalang street, Longhua District, Shenzhen

0755-29008612 29008442

info@king-med.com

Official account of Lontek

WeChat

Copyright:Shenzhen Lontek Electronic Technology Co., Ltd
Copyrigh©2020 56sys.com Shenzhen Lontek Electronic Technology Co., Ltd. all rights reserved
Website filing number: 粤ICP备14053113号 | Copyrigh©2020 56sys.com Shenzhen Lontek Electronic Technology Co., Ltd. all rights reserved